BRADFORD MICROASSAY
- Prepare a standard curve of five to six concentrations of
a protein standard, which is representative of the protein solution
to be tested. Usually 0 - 10 ug of protein adjusted to 800 uL
with water or sample buffer.
- Pipet 10 uL each sample into a tube and adjust volume to
800 uL (NOTE: the volume of the unknown can be adjusted to as
much as 800 uL). Samples are usually assayed in duplicate or
triplicate.
- Add 200 uL of the dye reagent (DO NOT DILUTE). Vortex each
tube.
- Incubate at room temperatre for at least 5 minutes. Absorbance
will increase over time; samples should incubate at room temperature
for no more than 1 hour.
- Measure absorbance at 595 nm